Sirem/Cartridge Tests; Sirem Input Verification; Monitor Shutdown Test - Siemens SIRECUST 1280 Service Manual

Bedside monitors
Table of Contents

Advertisement

E331.E2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
5.4.
b.
Press the INPUTS key. Verify
that the BED value changes
from a 1 level to a O level.
c.
Press the SYNC key again.
d.
Press the INPUTS key. Verify
that the BED state changes
from a 0 level to a 1 level.
SIREM/Cartridge Tests
CAUTION:
Switch the monitor power OFF
before connecting or disconnecting the
SIREM
cable
to
prevent
equipment
damage.
5.4.1.
1.
5.4.2.
Monitor Shutdown Test
Connect a SIREM equipped with a
LIM and any two of the following
cartridges
(if
available)
to
the
SIREM 1 input, X16, on the back
of the monitor.
* PCO2 Cartridge with transducer,
+ Sa02 Cartridge,
* NIBP Cartridge.
Rapidly
eject all cartridges from
the SIREM and reinsert them.
Does the monitor shutdown?
* Yes:
Go
to
Chapter
6,
Section 6.7F.
e No:
Switch
the
monitor
OFF
and
continue with test.
SIREM Input Verification
Connect a patient simulator to the
EKG and PRESSURE inputs of the
LIM cartridge.
Set
the
simulator
controls
to
provide an EKG waveform
of 80
BPM
©
1mV,
and
two
blood
pressure curves.
Insert Temperature Simulator Test
Plugs into the TEMP input jacks of'
the LIM cartridge.
Set
the
monitor
to
display
the
active
patient
input
parameters
provided by the simulator. Refer to
the Operating Instructions for the
installed software version,
Order
No. A91004-M3331-L079-XX-7600.
Is the display in accordance with
the
configuration
of the
SIREM
and as indicated in the Operating
Instructions.
* Yes:
Continue with test.
* No: Refer to Chapter 6, Section
6.7K.
Examine
each
waveform
for the
presence of noise spikes, glitches,
or voids.
Is the display free of noise and
voids?
* Yes:
Continue with test.
* No:
Go to Chapter 6, Section
6.7R.
Switch
monitor
power
OFF
Disconnect the SIREM cable from
X16, and connect it to X17.
Switch monitor power ON.
The
active
parameters
are
dis-
played on the monitor screen.
Is the display in accordance with
the
configuration
of the
SIREM
and as indicated in the Operating
Instructions?
.
* Yes:
Continue with test
+ No: Refer to Chapter 6, Section
6.7K.
Page 5-6