Maximum scatter radiation in the operating area

Scatter radiation in the main operating area according to EN 60601-1-3
Height
Measure-
above the
ment A
floor [cm]
[µGy/h]
10
20
30
40
1051
50
1760
60
2970
70
4342
80
4687
90
6394
100
8370
110
8960
120
8586
130
7938
140
7150
150
6232
160
5162
170
3604
180
2657
190
2030
200
1696
Tolerance of air kerma measurements ± 5%
Measurement A: Operating area A
Continuous fluoroscopy 110 kV, 2.9 mA, C-arm horizontal, downward beam direc-
tion without additional Cu filter, with scattered radiation grid
Measurement B: Operating area A
Characteristic HC2 high 68 kV, 1.3 mA, C-arm horizontal, downward beam direc-
tion without additional Cu filter, with scattered radiation grid
Measurement C: Operating area B
Continuous fluoroscopy 110 kV, 2.9 mA, C-arm horizontal, downward beam direc-
tion without additional Cu filter, with scattered radiation grid
Measurement D: Operating area B
Characteristic HC2 high 68 kV, 1.3 mA, C-arm horizontal, downward beam direc-
tion without additional Cu filter, with scattered radiation grid
Measure-
ment B
[µGy/h]
529
85
616
94
774
129
145
213
342
491
588
845
1245
1367
1353
1254
1057
765
555
397
314
257
220
Measure-
Measure-
ment C
ment D
[µGy/h]
[µGy/h]
37
4
42
4
43
4
49
5
75
9
118
15
147
20
201
26
244
28
270
32
283
34
285
37
287
37
285
39
284
37
283
37
288
39
298
40
298
40
298
40
Operator Manual
SPR2-310.620.30.02.02