ARCADIS Varic
SPR2-310.620.30.02.02
Configuring the operating programs
An examination set consists of a set of operating programs. Each of these pro-
grams defines the settings for X-ray generation and image acquisition for one
single operating mode.
The operating programs are configured in the Operating Program
Configuration dialog, which you have called up from the Examination Set
Configuration dialog.
The Operating Program Configuration dialog contains a tab card (labeled corre-
spondingly) for each supported operating mode. In the tab cards you can specify
the parameters of the operating program in the respective operating mode.
In addition, some parameters may be common to several operating modes of an
examination set (e.g. edge enhancement). They are available on other tab cards
that are independent of the operating mode.
The parameters for the Fluoroscopy, Pulsed Fluoroscopy and Digital Radiography
modes are configured for every examination set. For the Subtraction option the
parameters for the Subtraction and Roadmap modes are configured additionally.
The parameters are individually adapted to each examination set.
There is a Default button on every tab card. With this button you can reset the
parameters back to the original state (parameter settings of the underlying oper-
ating program).
Configuration
0
13
of 24