SPR2-310.620.30.02.02

Moving the C-arm orbitally

Starting from the basic position (0°), you can swivel the C-arm by up to +90° or
up to -40° (130° in total).
Release the brake marked in blue (arrow).
Swivel the C-arm to the required orbital position while observing the blue
scale.
Lock the brake again.
130°
System Description
130°