Guidelines And Manufacturer's Declaration - Electromagnetic Interference Immunity - Siemens ARCADIS Varic Operator's Manual

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

Interference immunity
tests
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient dis-
turbances/bursts
according to IEC 61000-4-4
Surge
according to IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
according to IEC 61000-4-11
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
according to IEC 61000-4-11
Note: U
is the AC mains voltage before application of the test level.
T
ARCADIS Varic
SPR2-310.620.30.02.02
Guidelines and manufacturer's declaration –
Electromagnetic interference immunity
The system is intended for use in an electromagnetic environment as specified
below. The customer or the user of the system should ensure that it is operated
in such an environment.
IEC 60601
Compliance level
test level
± 6 kV contact dis-
± 6 kV contact dis-
charge
charge
± 8 kV air dis-
± 8 kV air dis-
charge
charge
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input/
± 1 kV for input/
output lines
output lines
± 1 kV differential-
± 1 kV differential-
mode voltage
mode voltage
± 2 kV common-
± 2 kV common-
mode voltage
mode voltage
< 5% U
*
not possible
T
(> 95% dip in U
)
T
for 0.5 cycles
40% U
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
< 5% U
< 5% U
T
(> 95% dip in U
)
(> 95% dip in U
T
for 5 seconds
for 5 seconds
Electromagnetic environment - guidance
Floors should be made of wood, concrete or
ceramic tiles. If the floor is covered with synthetic
materials, the relative humidity must be at
least 30%.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Compliance with the low values for the leakage cur-
rent requires the use of inductances. The resultant
necessary remagnetization of these inductances
causes in the most unfavorable case a behavior
deviating from the requirements specified in
IEC 60601-1-2.
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the
system requires continued operation during power
mains interruptions, it is recommended that the
system be powered from an uninterruptible power
supply.
T
)
T
T
)
T
T
)
T
Technical Data
0
13
of 34