Switching The System Off; Emergency Shutdown Button (Installed On-Site); Standby Power Supply; Standby Power Supply Inside The Hospital - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

System Operation
On-Off/Emergency Stop
6
/ 32

Switching the system off

◆ Press the button on the generator ON/OFF console.
– The entire system and all further connected devices are immediately
switched off.
– All started operating sequences will be interrupted and the selected
programs will be deselected.
➩ Refer to the FLUOROSPOT Compact Operator Manual for further information.
Emergency SHUTDOWN button (installed
on-site)
Use this shutdown method only in an extreme emergency because it is an un-
controlled process!
Data could be lost, e.g. unsaved images, exporting and filming jobs, etc. The tube
cooling system is also disconnected from the power supply, and the tube can
overheat.

Standby power supply

Only continuous fluoroscopy is possible in standby operation.
Pulsed fluoroscopy and all radiography modes are not possible.
"Standby power supply active" appears on the integrated generator control con-
sole in the message line.

Standby power supply inside the hospital

A standby power supply (inside the hospital) switches on if the line voltage fails.
However, this is not as a rule without interruption and can take some time.
The system is switched off and must be switched on anew.

Battery-supported standby power supply (optional)

UPS = Uninterrupted Power Supply
An uninterrupted power supply is a battery-supported system for the standby
power supply in the case of mains power supply failures.
If an UPS is installed for your system, then this UPS switches automatically to
standby operation if the mains voltage fails or is too low. To prevent loss of data/
images, the imaging system remains switched on. The generator and table are
switched off.
AXD3-340.620.13.01.02
Operator Manual