Configuration
Parameter settings
20
of 24
The Roadmap tab card
This tab card is available only with the Subtraction option.
You can change the following parameters on the tab card:
Parameters
Opacification
Dose Level
Landmark
Bone Display
Autostore
Storage Rate f/s
Autoloop
Parameter Set
You have to select the setting "black" for the Bone Display parameter. Other-
wise, the system will apply incorrect image brightness values and automatic
vessel contour detection fails.
Remarks
Possible values: min, max
Possible values: medium, high
With the slider: 0 to 30%, in steps of 5%
Possible values: white, black
off (no automatic storage)
all (all images are stored)
Only if Autostore = all
Setting in frames/s
Check box for activating automatic play-
back of a sequence of images at the end
of the exposure
Default parameter set for automatic ves-
sel contour detection
SPR2-310.620.30.02.02