Image quality

The attachment of a dedicated option must not affect image quality (impairment
of the diagnosis).
After maintenance or service work, the correct function of the non-Siemens sys-
tem on the image intensifier must be tested.

Electrical safety

EN 60601-1, Section 3, "Protection against electric shock hazard" must be com-
plied with.

Electromagnetic compatibility

EN 60601-1-2 must be observed in order to comply with the limit values for elec-
tromagnetic compatibility.

Additional safety information

Risk of injury due to sharp edges must be avoided.
To avoid thermal overloading of components and short circuits, EN 60601-1
Section 7 and, if appropriate, UL 60601-1 must be complied with.
Connecting external loads to the power supply of the C-arm system is not per-
mitted.
We recommend that users in the EU have the relevant manufacturer of the
accessory operated by you, e.g. a navigation system, confirm the CE Declara-
tion of Conformity according to Appendix II, MDD and the Declaration of
Compatibility according to Article 12, MDD. In countries outside the EU the
relevant national regulations must be observed.
Product liability and warranty are restricted or expire if the above listed conditions
and limit values are not complied with when attaching accessories, such as nav-
igation systems.
For non-Siemens options, such as navigation systems, we generally accept no
liability.
Safety