Switching to emergency power supply

In the event of a power interruption of longer than 8 ms, the ARCADIS Varic can
switch off. In this case the ARCADIS Varic must be switched on again after the
system has switched over to the emergency power supply.
In case of a power failure, a signal sounds (up to 10 min.) when the system
switches to uninterruptible power supply (UPS).

Pulling out the power supply plug

After the power plug is disconnected, the uninterruptible power supply (UPS)
supplies voltage to the imaging system and the left-hand monitor until the
ARCADIS Varic switches off completely.
When the power plug is pulled out, switching to the uninterruptible power supply
causes an acoustic signal to be emitted. The UPS switches off after 10 min.
at the latest.
As soon as the mains supply is restored, the battery of the UPS is recharged.
Please remember that the UPS battery life is limited.

Emergency STOP

Immediately press the red EMERGENCY STOP button (arrow) on the electronics
unit of the C-arm system at the first sign of any dangerous situation resulting from
motorized movements.
S T O P
Motorized vertical movement is then immediately disabled.
All other system functions remain unaffected by this.
Unlock the button only after the danger has clearly been eliminated.
The button can be unlocked by gently turning it clockwise.
Safety
0
0
0
11