Current/voltage values
Voltage
100 V
110 V
120 V
127 V
200 V
230 V
240 V
Pulsed Fluoroscopy
Fluoroscopy
Digital Radiography
Current-time-reference
product
Tolerances
Power rating
Exposure times
ARCADIS Varic
SPR2-310.620.30.02.02
Long-term current consumption
17 A
16 A
15 A
14 A
9 A
8 A
8 A

Generation of radiation

Generator

Min. pulse width 7 ms
Pulse rate 15 f/s/23 mA
Max. pulsed output 2.3 kW
40 kV to 110 kV/0.2 to 15.2 mA (max. 1000 W)
40 kV to 110 kV/0.2 to 23 mA (max. 1000 W)
1.6 mAs/102 kV
kV ± 10% (measured with spectrometric kV method)
mA ± 8% ± 0.1 mA (measured in rectified high-voltage circuit)
Fluoroscopic time 1 digit (6 s) ± 5%
1.4 kW (102 kV/1.6 mAs)
Min. pulse width 7 ms
Max. pulsed output 2.3 kW
min. 50 ms at 40 kV 20 mA
max. 10 s
Short-term current consumption
29 A
26 A
24 A
23 A
15 A
13 A
12 A
Technical Data