ARCADIS Varic Switching on

The ARCADIS Varic is operated via a grounded wall socket. The mains cable is on
the monitor trolley.
Plug the power plug into the appropriate socket.
– Power connection is now established.
Position the unit so that it is easy to disconnect from the mains.
Warning
Contact voltage while using the unit.
Electrical shock to operator and patient
Connect this device only to a power supply network that has been properly
grounded.
Press the ON switch at the monitor trolley.
– The ARCADIS Varic is switched on.
– The system automatically runs a self-test.
Depending on which shutdown procedure of the ARCADIS Varic was used
beforehand, the system will be ready for operation again in less than a minute
(following a simple shutdown) or after 3 min (following a complete shutdown)
(
Page 34).
If the HIPAA Security Package is activated, it can take approx. 90-100 s for the
ARCADIS Varic to be operational again after a simple shutdown.
If the ARCADIS Varic is restarted after a simple shutdown and immediately after-
wards large image data amounts (500 MB and more; approx. 250 single images)
are loaded, an error message may be displayed.
In this case, confirm the error message and repeat loading the data.