Section II

Connection to the Electric Supply

It is recommended to connect the pump to the power supply through a dedicated
CAUTION!
safety switch on the main electrical panel of the installation, or in proximity of
the power supply connection point.

Inlet and exhaust connections

Remove the protective caps from both ports. Connect the system to
be evacuated to the inlet flange, using a centering ring with OR and a
clamp flange.
For guaranteed reliable sealing, use an OR gasket in FKM.
NOTE
The inlet duct is equipped with a sieve filter preventing solid parts
from entering and damaging the pump.
When the gases to be pumped out contain dust, it is advisable to insert a dust
NOTE
filter before the inlet flange.
NOTE
When the gases to be pumped out contain large quantities of vapor, it is
advisable to include a condense separator before the inlet flange.
To make best use of the pump's capacity, use only short, straight
piping, with a diameter not smaller than that of the inlet flange.
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (H.00)