Ac Current (Sine Wave) Specifications - Fluke 5520A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

1-18. AC Current (Sine Wave) Specifications

Range
Frequency
29.00 µA to
10 Hz to 20 Hz
329.99 µA
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30
kHz
0.33 mA to
10 Hz to 20 Hz
3.2999 mA
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30
kHz
3.3 mA to
10 Hz to 20 Hz
32.999 mA
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30
kHz
33 mA to
10 Hz to 20 Hz
329.99 mA
20 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
10 kHz to 30
kHz
0.33 A to
10 Hz to 45 Hz
1.09999 A
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
1.1 A to
10 Hz to 45 Hz
2.99999 A
45 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
3 A to
45 Hz to 100 Hz
10.9999 A
100 kHz to 1
kHz
1 kHz to 5 kHz
11A to
45 Hz to 100 Hz
20.5 A
100 Hz to 1 kHz
[2]
1 kHz to 5 kHz
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9% of
output + floor as shown.
[2] Duty Cycle: Currents < 11 A may be provided continuously. For currents > 11 A, see Figure 1-4. The
current may be provided 60-T-I minutes any 60 minute period where T is the temperature in °C
(room
temperature is about 23°C) and I is the output current in Amps. For example, 17 A, at 23°C could be
provided for 60-17-23 = 20 minutes each hour.
[3] For compliance voltages greater than 1 V, add 1 mA/V to the floor specification from 1 kHz to 5 kHz.
[4] For compliance voltages greater than 1 V, add 5 mA/V to the floor specification from 5 kHz to 10 kHz.
LCOMP off
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µA)
90 days
1 year
0.16 + 0.1
0.2 + 0.1
0.12 + 0.1
0.15 + 0.1
0.1 + 0.1
0.125 + 0.1
0.25 + 0.15
0.3 + 0.15
0.6 + 0.2
0.8 + 0.2
1.2 + 0.4
1.6 + 0.4
0.16 + 0.15
0.2 + 0.15
0.1 + 0.15
0.125 + 0.15
0.08 + 0.15
0.1+ 0.15
0.16 + 0.2
0.2 + 0.2
0.4 + 0.3
0.5 + 0.3
0.8 + 0.6
1.0 + 0.6
0.15 + 2
0.18 + 2
0.075 + 2
0.09 + 2
0.035 + 2
0.04 + 2
0.065 + 2
0.08 + 2
0.16 + 3
0.2 + 3
0.32 + 4
0.4 + 4
0.15 +20
0.18 + 20
0.075 + 20
0.09 + 20
0.035 + 20
0.04 + 20
0.08 + 50
0.10 + 50
0.16 + 100
0.2 + 100
0.32 + 200
0.4 + 200
0.15 + 100
0.18 + 100
0.036 + 100
0.05 + 100
0.5 + 1000
0.6 + 1000
2.0 + 5000
2.5 + 5000
0.15 + 100
0.18 + 100
0.05 + 100
0.06 + 100
0.5 + 1000
0.6 + 1000
2.0 + 5000
2.5 + 5000
0.05 + 2000
0.06 + 2000
0.08 + 2000
0.10 + 2000
2.5 + 2000
3.0 + 2000
0.1 + 5000
0.12 + 5000
0.13 + 5000
0.15 + 5000
2.5 + 5000
3.0 + 5000
Introduction and Specifications
Max
Distortion &
Noise 10 Hz
Compliance
to 100 kHz
adder
BW
± (µA/V)
± (% output
+
floor)
0.15 + 0.5 µA
0.05
0.1 + 0.5 µA
0.05
0.05 + 0.5 µA
0.05
0.5 + 0.5 µA
1.5
1.0 + 0.5 µA
1.5
1.2 + 0.5 µA
10
0.15 + 1.5 µA
0.05
0.06 + 1.5 µA
0.05
0.02 + 1.5 µA
0.05
0.5 + 1.5 µA
1.5
1.0 + 1.5 µA
1.5
1.2 + 0.5 µA
10
0.15 + 5 µA
0.05
0.05 + 5 µA
0.05
0.07 + 5 µA
0.05
0.3 + 5 µA
1.5
0.7 + 5 µA
1.5
1.0 + 0.5 µA
10
0.15 + 50 µA
0.05
0.05 + 50 µA
0.05
0.02 + 50 µA
0.05
0.03 + 50 µA
1.5
0.1 + 50 µA
1.5
0.6 + 50 µA
10
0.2 + 500 µA
0.07 + 500 µA
1 + 500 µA
[3]
2 + 500 µA
[4]
0.2 + 500 µA
0.07 + 500 µA
1 + 500 µA
[3]
2 + 500 µA
[4]
0.2 + 3 mA
0.1 + 3 mA
0.8 + 3 mA
0.2 + 3 mA
0.1 + 3 mA
0.8 + 3 mA
1
Specifications
Max
Inductive
Load
µH
200
200
50
50
2.5
2.5
1
1
1-17