Electrical Measurement Specifications
The accuracy specification is defined as ± (% reading +digit counts) at 23 °C ± 5 °C, ≥ 80% RH. Between -10 °C and 18 °C
and between 28 °C and 40 °C, accuracy specifications may degrade by 0,1 x (accuracy specification) per °C.
AC Voltage Measurement (V)
Range
500 V
Continuity Testing (R
LO
Range
(Autoranging)
20 Ω
200 Ω
2000 Ω
The test current is 200 mA for resistances of 2 Ω or less. The number of possible continuity tests with a fresh set of
batteries is 3000.
Test Probe Zeroing
Resolution
0,1 V
)
Resolution
0,01 Ω
0,1 Ω
1 Ω
Accuracy
Input Impedance
50Hz - 60Hz
0,8% + 3
Open Circuit Voltage
> 4 V
•
Press the Z button to zero the test probe.
•
Can subtract up to 2 Ω of lead resistance.
•
Error message for >2 Ω.
Electrical Installation Tester
Overload Protection
3,3 MΩ
660 V rms
Accuracy
+ (1,5%+3 digits)
+ (1,5%+3 digits)
+ (1,5%+3 digits)
Specifications
47