Slide
Slide
group box
Slide Adapter Format
Slide Adapter
Format
group box
2004-11
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
6. LSServer, Scanner Shell and Download
The Slide group box contains the following elements:
Slide Thickness box:
In the Slide Thickness box the thickness of a sample (i.e. a slide)
must be entered in order to allow the instrument to search for the focal
plane within an appropriate range.
If the Slide Thickness is entered incorrectly, the algorithm might be
unable to find the right surface.
The Slide Thickness ranges from 0 to 2000 µm and must be adjusted
respectively to the Adapter height defined in *.saf files (see Advanced
Menu - Edit SAF... on page 6-85):
1) If the Adapter height is measured from the surface of the plate
transport to the surface of the spots, the Slide Thickness must
be set to 0.
2) If the Standard Tecan Slide Adapter is used (with the default
height of 11.5 mm), the thickness of the slide that is inserted
into the adapter must be edited as well.
The Slide Adapter Format group box contains the following elements:
Load button
Click Load and the Open Slide Adapter Format dialog box appears.
Select a Slide Adapter Format (*.saf) file from the drop-down list and it
will be loaded automatically.
Load Standard Adapter button
Click the Load Standard Adapter button to load the Standard.saf file
and to reset default Geometry Parameters.
6-33