Replacing The Fuse - tecan ls series Operating Manual

Table of Contents

Advertisement

8. Cleaning & Maintenance
8.3

Replacing the Fuse

8-2
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
Turn the instrument OFF and unplug the main power cable from the
instrument. The fuse is located to the right of the main power cable
connection in the rear of the instrument.
Open the fuse compartment by inserting a screwdriver into the slot in the top
of the cover and push the fuse holder out.
Main power socket
Pull the fuse out and replace the defective fuse(s) with the spare fuse(s).
Ensure that the fuse(s) has/have the correct rating: T 2.0 A / 250 V (slow
blow)
OF THE SAME TYPE AND RATING.
Replace the fuse holder.
Reconnect the mains cable.
Turn the instrument on.
If the fuse continues to blow, call for service.
On/Off Switch
1 = ON
0 = OFF
WARNING
RISK OF FIRE!
REPLACE ONLY WITH FUSES
Caution
Fuse compartment
2004-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ls300

Table of Contents