6. LSServer, Scanner Shell and Download
6-72
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
The Define current loaded filter slides dialog box represents all filter slides
found and defined in the currently connected instrument. For each slide that is
loaded a different tab-dialog box is displayed. When connected to a Tecan LS
SERIES SCANNER instrument, two tabs Filter slide 1 (filter in channel 1 slot)
and Filter slide 2 (filter in channel 2 slot) are shown.
The Wavelength , Bandwidth and a Name for the specific dye used for detection
can be defined. Click OK to save the filter definition changes and close the dialog
box.
Each filter slide contains 6 filter positions:
Position 1 is reserved for focus measurement.
Positions 2 - 5 are free for filter definition.
Position 6 is reserved.
Position 1 is always reserved for focus measurement and cannot be
Position 6 is reserved for future functions and cannot be changed.
Empty filter slide positions (Wavelength / Bandwidth) must be set to 0.
No spaces are allowed in the filter name.
The Define current loaded filter slides dialog box is divided into individual tabs:
Filter Slide 1 tab
Filter Slide 2 tab
The Define current loaded filter slides dialog box contains the following
buttons:
OK
Click the OK button to save the selected options and to close the
dialog box.
Click the Cancel button to close the dialog box. The selected options
Cancel
are not saved.
Help
Click the Help button to display the online help for the dialog box.
Note
changed.
Note
Note