In the Setup menu, select Setup Port... .
This option allows an instrument to be connected. The Setup Port dialog box is
displayed, where the desired instrument and communication port can be
selected.
For more information refer to the chapter 6.2 Tecan LS Scanner Interface
description in this manual.
In the Setup menu, select Define Filter Slides...
The Define Filter Slides option is used to define the filter slides found in the
currently connected instrument.
For example in Dual Channel Instruments, two filter slides can be loaded
simultaneously. Parameters and comments for one / two filter slides (channel 1 /
channel 2) must be set before starting measurement (The first time is done by a
service engineer):
1. Load the selected filter(s) into channel 1 slot by hand.
You cannot define a filter slide that has been loaded into the channel 2
slot! All filters must be defined in channel 1 slot.
2.
In the Setup menu select Define Filter Slides... The Define current loaded
filter slides dialog box is displayed.
KEEP IN MIND: THIS OPTION IS NOT FOR GENERAL USE!
The instrument can recognize predefined filter slides and the filter
values should not be changed. However, if the filter in the filter slide
have been changed (by a service engineer) or if a new undefined
customized filter slide is to be used, the filter slides need to be defined.