tecan ls series Operating Manual page 127

Table of Contents

Advertisement

The Reset Transport Lock window
2004-11
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
Instrument is initialized and Tecan Scanner Control is now available .
Save packing materials and transport locks
for further transportation purposes. The LS series scanners must be
shipped with the original packing
and the transport locks must be installed.
BEFORE TRANSPORTATION
Place the instrument carefully in the indicated orientation before fixing the
transportation locks, otherwise the instrument will become damaged.
In the Reset Transport Lock window lock status is removed from the instrument,
transportation lock is reset and transports are initialized.
Click Next to start the removal of the lock status and to initialize the transports.
A Status Control bitmap shows the result of the operation.
The Reset Transport Lock window:
The Next button guides the user to the next screen of the wizard: the End
Remove Transport Lock window.
6. LSServer, Scanner Shell and Download
Caution
Do not try to open or close
the instrument flap manually.
Caution
CAUTION
6-97