tecan ls series Operating Manual page 106

Table of Contents

Advertisement

6. LSServer, Scanner Shell and Download
Demo Mode Allowed
Transport Lock...
6-76
Operating Manual for the LS Series Scanner No. 30001798 Rev. No. 1.5
The Test Command dialog box contains the following elements:
Command combo box
Answer edit box
Empty command input
field after sending
check box
Send button
Cancel button
To enable the Demo Mode option button in the Setup Port dialog box , select
Demo Mode Allowed from the Service menu.
In Demo Mode the Tecan LSServer can be simulated, no instrument needs to
be connected. This option is used for training purposes.
For further information refer to the chapter 6.2 Tecan LS Scanner Interface
description in this manual.
To start the Transport Lock wizard, select Transport Lock... from the Service
menu. A wizard guides the user through the transport lock procedure..
The Transport Lock option is implemented in Scanner Shell as a wizard. It
represents workflow modules, which define a sequential procedure and guide the
user step-by-step to perform a specific complex activity.
For transportation purposes it is recommended to lock the connected instrument.
A transportation shock absorber pad is inserted between the plate transport and
the optic chassis to avoid damage to the optical parts.
For further information refer to the chapter 6.4.9 The Transport Lock Wizard
description in this manual.
During the Transport Lock procedure:
1. The instrument door is opened.
2. The plate transport is moved into the correct position.
3. The user is asked to insert the transportation shock absorber pad.
4. Instrument is locked.
5. The instrument door is closed.
As long as the instrument is locked, the LSServer Measurement Control cannot
be started. Thus scanning is not possible.
For information unlocking instrument see 6.4.10 The Transport Unlock Wizard
description in this manual.
Use the Command combo box edit field to type in the
command that should be sent to the instrument.
Each command entered is automatically added to the
Command combo list: All commands sent can be
repeated by clicking on the corresponding list item.
The Answer edit box shows answers received from
connected instrument.
Selecting this box deletes commands typed in
Command combo box after they have been sent to
the instrument. New commands can be entered
immediately.
If the same command needs to be repeated several
times this box should not be selected.
Click the Send button to send the command, which is
in Command combo box edit field, to the connected
instrument.
Click the Cancel button to close the dialog box.
2004-11