4.2.6 MEASUREMENTS Screens
T
he i/iX Series uses
the output of the Source. This data acquisition system digitizes the voltage and current waveforms and
calculates several parameters from this digitized data. The results of these calculations are displayed
in a series o
f m
this information.
The Measurement screen
anywhere. Instead, these
be reached by successive
that for i Series models, on
measurement screens are available.
The
PHASE key ma
selected phase will be indicated by the phase LED's to
a
re selected,
data is only s
The following parameters
ENTRY
VOLTAGE
CURRENT
FREQ
TRUE POWER
VA POWER
POWER FACTOR
PEAK CURR
California Instruments
a DS
P based data acquisition system to provide extensive informatio
easurement data screens

Figure 4-7: MEASUREMENT Screen

s available on the iX Series are not menus in that no changes can
screens provide load parameter readouts. The measurement screens
ly pressing the MEAS key, which will toggle to all available screens. N
ly the first two screens are available. For the iX
y be u
sed to toggle through each output phase's measurement screens. Th
hown for phase A.
are available in the measurement screens:
DESCRIPTION
This value is the true rms output voltage measured at the voltage
sense lines.
This value is the true rms output current drawn by the load.
The output frequency is measured at the sense lines.
This value is the real power.
This value is the apparent power.
This readout shows the power factor of the load.
This readout reflects the highest peak current value detected at the
output. This is a track and hold peak current measurement. To
measure inrush current for a unit under test, open the output relay
and reset the peak current value using the BACK key. Then program
the output voltage and frequency and turn on the output relay. The
peak current measurement will continuously track the maximum
current value detected until reset. See also PEAK CURRENT in
MEASUREMENTS 3 screen.
. A total of three measurement screens a
the left of the LCD display. If all three phases
MEASUREMENTS 1
MEASUREMENTS 2
2253i / 2253iX
n regarding
re used to display all
be made
can
ote
series, all three
e
64

Advertisement

Table of Contents
loading

This manual is also suitable for:

2253ix

Table of Contents