California Instruments 2253i User Manual page 103

Ac and dc power source & analyzer
Hide thumbs Also See for 2253i:
Table of Contents

Advertisement

User Manual
7.6.2 P OWER BOARD
7.6.2.1 HIGH VOLTAGE DC INPUT
PINS
E1
E2
E3
7.6.2.2
HIGH VOLTAGE
PINS
E
4
E
5
E6
7.6.3 CONFIGURATION
The module may be configured to operate as two independent 375 VA LO RANGE half bridges or as
a single 750 VA dual ran
each power module, con
Note: Unless a module was exchanged in the field, the i/iX comes factory configured for the
correct mode of operation and these dip-switch settings should normally not have to
be changed. This information is provided for reference only.
Only factory authorized personnel should use this information if needed.
7.6.3.1 DIP SWITCH SETTINGS
DIP SWITCH S1
DIP switch S1 selects which oscillator phase will drive each half bridge.
DIP SWITCH
POSITION
1
2
3
4
5
6
DIP SWITCH S2 – 2 Positions
For 6005-701-1 control board (Assy rev H or higher) DIP switch S2 configures the source of the error
amplifier drive signal for single or three phase mode.
DIP SWITCH
POSITION
1
California Instruments
NAME
DESCRIPTION
+250V_IN
+250 Vdc @ <5 Adc, must be able to s
HV_COM_IN
Return for +/- 250V_IN
-250V_IN
-250 Vdc @ <5 Adc, must be able to sink current from module
OUTPUT
NAME
DESCRIPTION
HV_COM_OUT
Return for E5 and E6
OUTPUT_LEFT
0 – 150 Vac, 375 VA max from 120 Vac to 150 Vac, 3.13 Arms
max
OUT
PUT_
RIGHT
Same as for E5
ge half/full bridge. Because DIP switches are used to set the operation of
figuration may only be performed manually.
NAME
FUNCTION
LA
ON for LEFT = Phase A (LB, LC must be OFF)
RA
ON for RIGHT = Phase A (RB, RC must be OFF)
LB
ON for LEFT = Phase B (LA, LC must be OFF)
RB
ON for RIGHT = Phase B (RA, RC must be OFF)
LC
ON for LEFT = Phase C (LA, LB must be OFF)
RC
ON for RIGHT = Phase C (RA, RB must be OFF)
NAME
FUNCTION
IL
ON to connect error amplifier signal to the master error amplifier
signal. This switch must be ON for any single module system, or if
the module is a master in a multiple module system. This switch
must be OFF if the LEFT amplifier of the module is auxiliary to a
2253i / 2253iX
ink current from module
103