Figure 9-7: Power Interrupt For Group2/A (Nf) And Group3/A(Wf); Table 9-5: Normal Voltagesurge Sequence - California Instruments 2253i User Manual

Ac and dc power source & analyzer
Hide thumbs Also See for 2253i:
Table of Contents

Advertisement

User Manual
F1
T2
Test no.:
Standard:
T1
(ms)
F1 (Hz)
F2 (Hz)
Fmax =
65
0 Hz for G
Fmax = 800 Hz for Group3/A(WF)
T2 = 20 msec
T3 = 5 msec
VOLTAGE SURGE
This test requires 160V output voltage. If the power source is set at the low voltage range, the high
voltage range will be selected before the test starts. At the end of the test, the power source will be
switched back to the low range automatically.
Seq. No.
1
2
3
4
5
The output voltage will follow the sequence in Table 9-5. The above sequence will repeat itself three
times. Each repeat will start from sequence two. RTCA and Group 1 will run at 400 Hz. Group 2 and
A(NF) will run at 360 Hz and 650 Hz. Group 3 and A(WF) will run at 360Hz and 800Hz. The
frequency will return to the nominal setting when the test is completed. The BACK key (backspace)
will terminate the test at any time.
California Instruments
T1
0
Vol
t
21
22
I
II
5
0
50
360
Fmax
Fmax
360
roup
2/A(N
F)

Figure 9-7: Power Interrupt for Group2/A (NF) and Group3/A(WF)

RTCA
Group 1
115
115
160
160
115
115
60
70
115
115

Table 9-5: Normal VoltageSurge Sequence

F2
T3
23
24
III
IV
100
10
0
360
Fmax
Fmax
360
Voltage
Group 2
115
160
115
70
115
25
26
V
VI
200
200
360
Fmax
Fmax
360
Time
Group 3
ALL
115
5 Minute
170
30msec
115
5 Sec.
70
30msec
115
5 Sec.
2253i / 2253iX