Testing Continuity; Checking Diodes - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

Shift
Shift
34405A User's and Service Guide
Downloaded from
www.Manualslib.com

Testing Continuity

• Measurement Method: 0.83 mA ± 0.2% constant current source, < 5 V open circuit voltage.
• Response Time: 70 samples/ second with audible tone
• Continuity Threshold: 10 Ω fixed
• Input Protection: 1000 V (HI terminal)
Open Circuit Display:
Cont )
)
)
Typical Closed Circuit Display:

Checking Diodes

• Measurement Method: Uses 0.83 mA ± 0.2% constant current source, < 5 V open circuit voltage.
• Response Time: 70 samples/ second with audible tone
• Input Protection: 1000 V (HI terminal)
Reverse Bias or Open Diode Display:
Freq
Freq
Typical Forward Biased Diode Display:
manuals search engine
12A
Fused
1.25A/500V FH
Fused
1.25A/500V FH
V DC
Getting Started Tutorial
V
HI
12A
1000VDC
Open or
rms
750VAC
LO
1.2A
500Vpk
rms
I
CAT II (300V)
12A
V
HI
12A
1000VDC
rms
750VAC
Forward Bias
LO
1.2A
500Vpk
rms
I
CAT II (300V)
1
Test
Current
Closed
Circuit
Test
Current
23

Advertisement

Table of Contents