Table 4-7 Ac Volts Verification Test; Ac Voltage Verification Test - Agilent Technologies U3401A User's Manual And Service Manual

41/2 digit dual display multimeter
Hide thumbs Also See for U3401A:
Table of Contents

Advertisement

4
Performance Test

AC Voltage Verification Test

Configuration: AC volts
1 Select the AC voltage function.
2 Select each range in the order shown in

Table 4-7 AC volts verification test

Set the calibrator output to 0 V before disconnecting it from the multimeter input
C A U T I O N
terminals.
66
indicated input voltage and frequency. Compare measurement results to
the appropriate test limits shown in
appropriate source settling).
Function
V RMS
AC voltage
50.00 mV
450.00 mV
0.5000 V
4.5000 V
5.000 V
45.000 V
50.00 V
450.00 V
75.0 V
675.0 V
Table 4-
Table 4-
Input frequency
Range
1 kHz
500 mV
1 kHz
500 mV
1 kHz
5 V
1 kHz
5 V
1 kHz
50 V
1 kHz
50 V
1 kHz
500 V
1 kHz
500 V
1 kHz
750 V
1 kHz
750 V
7. Provide the
7. (Be certain to allow for
Error from nomial
one year
±650 μV
±2.65 mV
±3.2 mV
±17.2 mV
±32 mV
±172 mV
±400 mV
±2.4 V
±1.8 V
±4.8 V
U3401A User's and Service Guide