Cleaning Immersion Oil

2.3
(1)
Wipe with petroleum benzine.
(2)
Finish off the cleaning with absolute alcohol (ethyl or methyl alcohol) after cleaning with petroleum benzene.
If petroleum benzene is not available
If petroleum benzene is unavailable, use methyl alcohol alone. However, typically wipe three or four times because the
detergency is weak.
2.4

Decontaminating the Product

For routine disinfection of this product, Nikon recommends using 70% medical alcohol.
Use of organic solvents on plastic parts may result in discoloration.
Cautions on disposal
If contact occurs between a sample and this product, determine whether the sample is hazardous. If the sample is hazardous,
follow the standard procedures for your facility.
3

Restoring Microscope Data to Factory Default

Various configuration data can be restored to factory default with buttons on the Ni-E.
Use this method to restore objective device information to factory default.
<Procedure>
Turn on the power while holding down the following three
buttons on the front of the main body.
 DISPLAY Previous button
 DISPLAY Next button
 Z-RESET button
When successful, the display shown on the right side appears
and data is restored to factory default.
Chapter 3
Maintenance and Storage
56
Power switch ON
Turning on the power switch while holding down
three buttons
F L A S H D a t a I n i t i a l i z i n g .
Restoration to factory default settings completed