Attach The Condenser; [18.1] Motorized Universal Condenser - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

(3)
Pull the blown fuse and remove it from the fuse
holder. Place a new fuse in the fuse holder.
(4)
Push in the fuse holder.
18

Attach the condenser.

18.1
Motorized universal condenser
Tool: Hex wrench (3 mm across flats)
Combination with the motorized
universal condenser
When the motorized universal condenser is attached
first, the motorized XY stage cannot be mounted.
Attach the motorized XY stage before mounting the
motorized universal condenser.
(1)
Remove the tightening screw at the bottom of the
condenser using a hex wrench (3 mm across flats).
(2)
If the front cover is attached to the elevating section,
remove it.
(3)
Hook the half dovetail of the motorized condenser
onto the half dovetail groove of the lower substage,
push it against the reference surface, and then
tighten the fixing screw at the top of the substage to
secure.
Chapter 1
Assembly
Removing the tightening screw (bottom of the
Attaching the Motorized universal condenser
23
Blown
New
fuse
fuse
AC
IN
~
100-240V
2A
50/60Hz
Replacing the fuse
AC IN
~
100-240V
2A 50/60Hz
Fuse
AC250V, T6.3AL
Pushing the fuse holder
condenser)
Motorized condenser
fixing screw
Half dovetail
groove
Half dovetail
Tightening
screw

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents