24
Replace the ND filter (optional).
Tool: Hex driver (2 mm across flats)
The ND filter in the ND filter cassette of the Ni-E main
body can be replaced as follows:
(1)
Check that the power switch for control box A is OFF.
(2)
Press all ND filter IN/OUT switches to the "IN" side.
(3)
Remove the rubber cap attached to the top of the
filter cassette cover. (it can be removed with your
finger)
(4)
Insert a hex driver in the hole with the rubber cap
removed and loosen the internal screw. (The screw
is drop-proof and not detached from the cover.)
(5)
Hold the ND filter IN/OUT switch and pull out the
cassette.
(6)
Hold the end of the filter holding frame and widen
slightly to remove the filter.
(7)
When attaching the filter, place the filter on the
holding frame at the position indicated with the arrow
and slide it in.
25

Connect the motorized unit cable.

Precautions for connecting a cable
Be sure to turn off the power to the microscope and peripheral units before connecting a cable.
Refer to the connector figure and connect cables correctly.
• The extension cable included in the control box A is required to connect the dia-illumination lamphouse and the
LAMP for the control box A.
• An USB cable is required to connect the USB (H) for DS-L3 and the USB for the microscope.
• The C-CTC Camera Trigger Cable U3/L3 is required to connect EXT.I/O connector for DS-L3/U3 with the DSC1
connector for the microscope or the DSC2 connector for the connector box.
• The NI-SHCS Motorized Shutter Cable Short is required to connect the motorized shutter with EPI SHUTTER or
DIA SHUTTER.
• The RS232C cross cable (commercially available, both connectors have a D-Sub9 pin (female) in shape) is
required to connect the RS-232C for the motorized HG precentered fiber illuminator with the HGFIE for the
microscope.
Precautions for connecting the camera trigger cable
When connecting the camera trigger cable with the DSC connector, be sure to insert it completely.