Attaching The Filter Cube - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

■ Attaching the filter cube

Precautions for attaching and removing the cube
• Be sure to check that the light source is turned off before attaching or removing the cube.
• Be sure to check that the power switch for control box A is turned off and then attach the cube by rotating the
cube switchover turret. Also for the motorized epi-fluorescence cube turret, attach it by manually rotating the
inside turret.
(1)
Pull out the filter cube replacement cover on the
front of the epi-fluorescence cube turret to remove it.
(2)
Attach the filter cube to the slot.
(3)
Insert the filter cube nameplate into the slot cover
window.
Insert into the same address window as the address
shown inside the slot.
(4)
Turn the cube switchover turret to attach the filter
cube in the remaining slots and also insert the
nameplate.
For bright-field microscopy
For bright-field microscopy, be sure that address 1 is
empty.
For motorized/intelligent epi-fluorescence cube
turret, only address 1 can be set to [OPEN].
(5)
Restore the slot cover back to its original position.
Chapter 1
Assembly
12
3
2
1
Filter cube
replacement slot cover
5
A
Address
indication
U V -2 A
EX 330
-38 0
DM 400
BA 420
Filter cube
A
Attaching the filter cube
5
4
Address
indication
Nameplate
window