Attaching The Dic Module To Dic Condenser - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

■ Attaching the DIC module to DIC condenser

D-CUO is provided with two sliders. Incorporate the oil condenser-specific DIC modules [D-C DIC N2 OIL] and
[D-C DIC NR OIL] to each slider.
Select the DIC module appropriate for the objective. (See the next table.)
To support for particular purposes, modules for higher contrast or resolution are available.
Note, however, that in principle the contrast contradicts the resolution of the differential interference contrast
image (the higher the contrast, the lower the resolution).
(1)
Attach the DIC module to the slider.
Identify the slider for the DIC module by the prism
name sticker ([N2] and [NR]) affixed to the handle,
and attach the DIC module to it.
Align the module orientation groove on the bottom
of the DIC module with the slider's orientation pin
and tighten the DIC module fixing screw on the
slider.
(2)
Insert the DIC module (slider) into the
condenser.
For the differential interference contrast
microscopy, insert the slider with the DIC module
until it reaches the condenser main unit.
Pull out the slider from the condenser when
switching over to the bright-field microscopy.
Chapter 1
Assembly
Slider
DIC module
name sticker
DIC module mounted:
29
Module orientation pin
DIC module
Module orientation
groove
DIC module fixing screw
Attaching the DIC module

Hide quick links: