Transportation (Using the fastening position mode switch)

When transporting the Ni-E main body or any motorized
accessories, the device must be fastened or moved into the
transportation position.
To fasten, press the fastening position mode switch at the rear of
control box A to "ON" side and perform the following:
Keep the fastening position mode switch OFF during normal use
of the microscope.
(1)
Check that the power switch for control box A is OFF. If it is
ON, turn it OFF.
(2)
Turn ON the fastening position mode switch.
(3)
Turn on the power for control box A.
This moves the motorized devices of the Ni-E and the
following motorized accessories attached to Ni-E to the
default fastening position.
 Motorized elevating section
 Motorized XY stage
 Motorized quadrocular tilting tube
 Motorized DSC Zooming Port
(4)
After moving, turn OFF the control box A power switch and
fasten the motorized elevating section, motorized XY stage,
and motorized quadrocular tilting tube with the anchoring
tool to transport. There is no anchoring tool for the
motorized DSC zooming port. Transport it in this fastening
position.
(5)
When fastened, turn OFF the fastening position mode
switch.
Devices fastened manually
Before transporting the following devices, be sure to fasten the device with the tightening screw or manually configure the
device so that it is in the stable state.
Motorized universal condenser
Attach the tightening screw removed from the bottom of the condenser in Chapter 1 "3 Assembly Method - 18.1 Motorized
universal condenser". Adjust the position of the condenser using a condenser centering screw and align the tightening
screw hole position with the attachment position in advance.
Manual quadrocular tilting tube
Attach the tightening screw removed from the bottom of the tube in Chapter 1 "3 Assembly Method - 12 Attach the
quadrocular tilting tube and DSC zooming port for quadrocular tube." Set the optical path switching dial to the [F] position
in advance.
Manual DSC zooming port
Set the zoom dial to 0.6x. (no tightening screw)
5

Storage

 Store this product in a dry location where mold is unlikely to form.
Storage conditions are as follows: temperature (-20°C to +60°C), humidity (90% RH max., no condensation)
 Store the objectives and eyepieces in a desiccator or similar container with a drying agent.
 Place a cover over this product to protect it from dust.
 Switch off the microscope (press the switch to the "O" position) and wait for the lamphouse to cool before covering this
product with a cover.
Chapter 3
Maintenance and Storage
57
Fastening position
mode switch
(ON: fastening position
mode,
OFF: normal mode)
OFF ON
FASTENING
N i - E
POSITION SW
CONTACT
ARM1
This device complies with Part
15 of the FCC Rules. Operation
is subject to the following
two conditions:
(1) this device may not cause
harmful interference, and
(2) this device must accept any
interference received,
including interference that
may cause undesired
operation.
This Class A digital apparatus
complies with Canadian ICES-003.
Cet appareil numérique de la
classe A est confirme à la
norme NMB-003 du Canada.
MODEL NI-CTLA
5 4 0 0 0 1
MADE IN JAPAN
100-240V~ 5A 50/60Hz
LAMP
Control box A (Rear)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents