15 Description of individual functions

15.1 ACAI function (automatic counting accuracy improvement = automatic
reference optimisation)
In order to improve add-up accuracy, the
ACAI function calculates the reference
weight anew each time new pieces are
added.
At every reference optimisation, the ref-
erence weight is calculated anew. As the
additional pieces increase the base for
the calculation, the reference also be-
comes more exact.
ACAI notes
• The ACAI procedure must be carried out after setting the reference weight. For
this purpose the piece number must still be on the weighing plate.
• Do not remove the piece number before the ACAI procedure is complete.
• If the pieces remain within the ADAI range, they must not be counted during add-
ing.
• If you want to have the most accurate add-up results for each different batch of
identical pieces, use the ADAI each time when starting a count for a new batch.
• The ACAI function is set by the factory to manual operation if the reference weight
is entered digitally via the keyboard, the ID memory or via PC via serial interface.
This can be converted to automatic mode. The ACAI mode during the input of the
reference weight via ID or digital input is controlled via function F-02-02 (see chpt.
14.2).

15.1.1 ACAI automatic mode

Setting in menu (chpt..14): F-02-01 = „1"
1. Run:
Leave the piece number on the weigh-
ing plate after setting the reference
weight
CWS-BA-e-0610
28
111.000
3.108
Kg
10
111.000
1.110
Kg
g
g