KERN CWS Operating Instruction page 42

Counting scale
Table of Contents

Advertisement

2. Add additional pieces within the next
ACAI range (see table below). As a
rough guide, approximately double the
piece number on the weighing plate.
Piece number on
the weighing plate
10
20
30
40
50
AUTO OPTIM.
As long as you are within
the range, the triangle
will pop up during adding.
3. Add more pieces within the ACAI range until you have reached a piece number
that is as great as the greatest piece number you wish to count.
4. When you have added the maximum number of required pieces, remove the
pieces and start the add-up procedure.
42
ACAI
Add-up range
13~26
23~49
33~70
43~89
53~106
AUTO OPTIM.
When adding is stopped
and the display becomes
stable, the triangle
start flashing.
2.331
Piece number on
the weighing plate
60
70
80
90
100
>200
Once the new reference
weight has been calcu-
will
lated, the triangle
disappear.
21
111.000
Kg
g
ACAI
Add-up range
63~122
73~138
83~152
93~166
103~299
203~492
AUTO OPTIM.
will
CWS-BA-e-0610