16.2 Data output

Parameters are set in function F-06-X (see chpt. 14).
Data output mode F- 06-01:
F- 06-01="0" .
Data are transmitted by pressing the PRINT key. Data transmission is indicated
by a flashing piece number display.
F- 06-01="1" .
Continuous data transmission The interval between continuous data is c. 2 sec.
F- 06-01="2" .
Data will be transmitted when the weight display is stable at +5d or above.
F- 06-01="3" .
Data is transmitted whilst the weight display is stable at ±5d or above/lower.
Selecting output values F- 06-02:
To select which data to transmit, enter either "0" or "1" for the data: ID no., piece
number, weight or reference weight
Example: When "1100" is entered only the ID no. and the piece number will be sent.
Data format F- 06-03:
Apply setting F- 06-03="2" to act as data format for general peripherals.
Baud rate F- 06-04:
F- 06-04="0" :2400 bps
F- 06-04="1" :4800 bps
F- 06-04="2" :9600 bps
48
8
F-06-02
88Clear8
8
ID No.
Quantity
Weight
Ref.weight
a11000
„0"= not send
„1"= send
CWS-BA-e-0610