Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is re-
quired for stabilization. Ensure that there are no objects on the weighing plate.
1. Press the PRINT key and keep on pressing it
when balance is turned off. To achieve this,
turn on the balance by using the ON/OFF key.
"CAL" appears on the display.
2. Press
button
3. Weight value of recommended adjustment
weight is flashing on display. However, you
may enter a value of your choice via the num-
ber keys. *
4. Press
button The display stops flashing.
5. Ensure that there are no objects on the weigh-
ing plate.
Press
button After successfully storing
the zero point "CAL F" will appear on the dis-
play.
6. Carefully place adjusting weight in the centre
of the weighing plate.
Press key
, adjustment will take place.
Once adjustment is complete, the display will
return to step 1. Remove the adjusting weight
and turn off the balance using the ON/OFF
key. Reattach the removed adjustment switch
cover.
In case of an adjustment error or an incorrect
adjusting weight, the display will not return to
step 1. Press the ON/OFF key and repeat the
adjustment procedure.
* The adjustment should be made with the recommended adjustment weight (see
chap. 1 "Technical data"). Weights of different nominal values may be used for cali-
bration but are not optimal for technical measuring.
Info about adjustment weights can be found on the Internet at: http://www.kern-
sohn.com
CWS-BA-e-0610
CAL
0.000
Kg
CAL 0
CAL 0
CAL F
CAL
6.000
Kg
6.000
Kg
6.000
Kg
6.000
Kg