7.5 Storage battery operation (optional)

Installation:
1. Disconnecting the power pack from the balance
2. Unscrew both M3 screws and remove the cover of the battery compartment.
3. Connect wires in the battery compartment to the battery
Connect red wire to positive (+/red) terminal and black wire to negative
(-/black) terminal. Otherwise there is a danger of explosion.
4. Insert battery in compartment and reattach the cover with the help of the
screws.
5. Turn on the balance and check whether the balance is working normally.
Battery
Screws
Battery compartment
Loading the battery
Use the provided power pack for battery charging.
Before the first use, the battery should be charged by connecting it to the mains
power supply for at least 15 hours.
• The display [Lo BAT] indicates that the capacity of the battery is nearly ex-
hausted. Connect the power adaptor as soon as possible to change the bat-
tery.
• Whilst working in battery mode, [
symbol
• It is possible to use the balance during charging
• Charge the battery at a temperature ranging between 5°C and 35°C
• Charge the battery at regular intervals (every 3-6 months), when the balance
is not put to use for a longer period of time.

7.6 Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To re-
ceive accurate measuring values it is also recommended to adjust the balance peri-
odically in weighing operation.
14
"+" electrode red
] will be displayed above the
"-" electrode black
battery