5-3 Adjusting the Judgement Parameters
Adjusting Judgement Parameters While Looking at Measurement Results
If correct judgements are not possible, you can move directly from the Setup Mode display to the judgement
parameters display to make adjustments.
[Test] – [Continuous test] – (Either display)
1
] − [Adjust judgement] on the right of the
Press [
display.
2
Adjust the judgement parameters.
Setting Up the Best Judgement Parameters Automatically
The judgement parameters of the selected inspection items can be automatically adjusted by using actual
workpieces which are considered as good and faulty products.
[Test] − [Continuous test]
1
Move to the inspection item for which you want to au-
tomatically adjust the judgement parameters and
press [
] – [Adjust judgement] on the right side of
the display.
2
Press [
] – [Auto adjustment].
3
Display a sample image of a good object and press
[OK Teach]. Display a sample image of a bad object
and press [NG Teach].
4
Repeat these steps for at least three samples each.
5
Press [Back].
The best judgement parameters will be set automatically.
6
Press [OK].
FQ User's Manual
Adjusting the Judgement Parameters
Upper limit
Lower limit
89