Omron Z302-E1-03 User Manual page 61

Vision sensor
Table of Contents

Advertisement

Step 2
Teaching
Teaching means to store the region and the edge position in the region as reference data for the measurement.
− [Inspection]
[Inspect]
1
Press [Teach].
2
Place the object that is to be used as the measure-
ment reference in front of the camera.
3
Move the rectangle to the location to be measured.
4
Press [OK].
5
Press [TEACH] on the lower right of the display.
The basic settings will be registered when teaching has
been completed.
6
Press [Back] to end teaching.
Changing the measurement region: p. 57
The following data is stored as basic measurement data.
Item
Reference data
Step 3
Adjusting Judgement Parameters
− [Inspection]
[Inspect]
1
Press [Judgement].
2
Set the range within which the D. Position is judged
as OK.
Continuous measurements will be performed for the im-
ages that are displayed.
3
Press [OK] to enter the value.
Item
Judgement Parameter
60
Inspecting with the Edge Position Inspection Item
− [Add item.] − [Edge Position] − [Settings] Tab Page
Parameter
Description
Reference position X
The reference coordinates (X, Y) of the position are set automatically.
Reference position Y
− [Add item.] − [Edge Position] − [Settings] Tab Page
Parameter
Setting
Range: −893 to 893
D. Position
Default: Lower limit: −893, Upper
limit: 893
The arrow in the middle shows the
direction for detecting an edge.
Drag to move
the region.
Changes the
width of the
Moves the
measure-
starting point of
ment region.
the measure-
ment region.
Lower limit
Upper limit
Blue for OK. Red for NG.
Description
Set the upper and lower limits from the
reference position as the offset within
which the judgement will be OK.
Moves the
end point of
the measure-
ment region.
FQ User's Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents