Scpi Programming; Table 3-1 Scpi Commands — Volts And Ohms Function - Keithley 6514 Instruction Manual

System electrometer
Hide thumbs Also See for 6514:
Table of Contents

Advertisement

While the electrometer is measuring ohms, the test current through the DUT and the voltage
drop across it can be displayed as follows:
V-Drop — While displaying an ohms reading, press SHIFT and then Ω to display the voltage
drop across the DUT. The "VΩ" message will indicate that a V-Drop reading is being displayed.
To return to the normal ohms reading, again press SHIFT and then Ω.
Test current — While displaying an ohms or V-Drop reading, press the Ω key. The test cur-
rent (ISRC) will be displayed for as long as you hold the key down.
WARNING

SCPI programming

Table 3-1
SCPI commands — volts and ohms function
Commands
[SENSe]
:FUNCtion <name>
:DATA?
:VOLTage
:GUARd <b>
:RESistance
:GUARd <b>
INITiate
READ?
A) SENSe:FUNCtion <name>
Parameters
Note that the parameter names are enclosed in single quotes ('). However, double quotes (")
can instead be used. Each measurement function "remembers" its own unique range setting.
The ohms function has a 250V compliance. To prevent electric shock,
always enable zero check to disable the test signal before making or break-
ing connections to DUT.
Description
SENSe Subystem:
Select function; 'VOLTage' or 'RESistance'.
Return latest "raw" reading.
Enable or disable guard.
Enable or disable guard.
Trigger one or more readings.
Trigger and return reading(s).
'VOLTage'
Volts function
'RESistance'
Ohms function
'CURRent'
Amps function
'CHARge'
Coulombs function
Volts and Ohms Measurements
Default
VOLT
3-7
Ref
A
B
OFF
C
OFF
C
B
B

Advertisement

Table of Contents
loading

Table of Contents