17-4
SCPI Reference Tables
Table 17-1 (cont.)
CALCulate command summary
Command
:SOURce2?
:STATe <b>
:STATe?
:FAIL?
:CLIMits
:CLEar
[:IMMediate]
:AUTO <b>
:AUTO?
:PASS
:SOURce2 <NDN> or
<NRf>
:SOURce2?
:NULL
:ACQuire
:OFFSet <NRf>
:OFFSet?
:STATe <b>
:STATe?
:DATA?
:LATest?
:CALCulate3
:FORMat <name>
:FORMat?
:DATA?
:SOURce2 <NDN> and <NRf> parameters:
<NDN> = #Bxxxx
= #Hx
= #Qxx
<NRf> = 0 to 15
Description
Query output pattern value.
Enable or disable limit 2 test.
Query state of limit 2 test.
Return result of limit 2 test; 0 (pass) or 1 (fail).
Composite Limits:
Clear I/O port and restore it back to
SOURce2:TTL settings:
Clears I/O port immediately.
When enabled, I/O port clears when :INITiate
sent.
Query state of auto-clear.
Define "pass" digital output pattern.
Specify 4-bit pass pattern (no failures).
Query "pass" output pattern.
Configure and control Rel:
Use input signal as Rel value.
Specify Rel value; -9.999999e20 to
9.999999e20.
Query Rel value.
Enable or disable Rel.
Query state of Rel.
Return all CALC2 readings triggered by INITiate.
Return only the last (latest) reading.
Path to configure and control CALC3 calculations
on buffer data:
Select buffer statistic; MEAN, SDEViation,
Maximum, MINimum or PKPK.
Query selected statistic.
Read the selected buffer statistic.
Binary format (each x = 1 or 0)
Hexadecimal format (x = 0 to F)
Octal format (x = 0 to 17)
Decimal format
Default
parameter Ref SCPI
OFF
ON
15
0.0
OFF
Sec 8
MEAN