17-6
SCPI Reference Tables
Table 17-4
SENSe command summary
Command
[:SENSe[1]]
:FUNCtion <name>
:FUNCtion?
:DATA
[:LATest]?
:VOLTage[:DC]
:NPLCycles <NRf>
:NPLCycles?
:RANGe
[:UPPer] <NRf>
[:UPPer]?
:AUTO <b>
:ULIMit <NRf>
:ULIMit?
:LLIMit <NRf>
:LLIMit?
:AUTO?
:GUARd <b>
:GUARd?
:XFEedback <b>
:XFEedback?
:CURRent[:DC]
:NPLCycles <NRf>
:NPLCycles?
:RANGe
[:UPPer] <NRf>
[:UPPer]?
:AUTO <b>
Description
Select function; 'VOLTage[:DC]',
'CURRent[:DC]', 'RESistance' or 'CHARge'.
Query measurement function.
Path to return instrument readings:
Return the last instrument reading.
Path to configure volts function:
Set integration rate in line cycles (PLC); 0.01 to
10.
Query NPLC.
Configure measurement range:
Select range; -210 to 210 (volts).
Query range value.
Enable or disable autorange.
Select autorange upper limit; -210 to 210
(volts).
Query upper limit for autorange.
Select autorange lower limit; -210 to 210
(volts).
Query lower limit for autorange.
Query state of autorange.
Enable or disable driven guard.
Query state of driven guard.
Enable or disable external feedback.
Query state of external feedback.
Path to configure amps function:
Set integration rate in line cycles (PLC); 0.01 to
10.
Query NPLC.
Configure measurement range:
Select range; -0.021 to 0.021 (amps).
Query range value.
Enable or disable autorange.
Default
parameter
Ref
SCPI
'VOLT'
Secs 3,
4, 5
Secs 3,
4, 5
6 (60Hz)
Sec 6
5 (50Hz)
Sec 6
21
(see Note)
210
2.1
OFF
Sec 3
OFF
Sec 11
6 (60Hz)
Sec 6
5 (50Hz)
Sec 6
2.1e-4
(see Note)