Table 18-7 Coulombs Measurement Accuracy Reading Limits - Keithley 6514 Instruction Manual

System electrometer
Hide thumbs Also See for 6514:
Table of Contents

Advertisement

7.
Table 18-7
Coulombs measurement accuracy reading limits
Nominal
Model
calibrator
6514 range
voltage
20nC
20V
200nC
200V
2µC
20V
20µC
200V
1
Nominal values.
2
Calculate actual voltage from applied charge and actual capacitance value: V = Q/C.
Verify charge measurement accuracy for each of the values listed in Table 18-7. For each
test point:
Select the correct Model 6514 measurement range.
Make connections to the correct standard capacitor.
Calculate the required voltage from the desired charge and actual standard capaci-
tance value: V = Q/C.
Disable zero check.
Set the calibrator voltage to the calculated value.
Verify that Model 6514 charge reading is within the required limits.
Enable zero check to discharge the capacitor.
Standard
1
capacitor
Applied charge
1nF
20.0000nC
1nF
200.000nC
100nF
2.00000µC
100nF
20.0000µC
Performance Verification
Actual
Model 6514 coulombs reading
2
voltage
limits (1 Year, 18˚C-28˚C)
_______V
19.9915 to 20.0805nC
_______V
199.195 to 200.805nC
_______V
1.97995 to 202.02005µC
_______V
19.7995 to 20.2005µC
18-19