9-8
Triggering
NOTE
DELAY Menu — To access the menu for trigger delay, press the DELAY key.
TRIG-OUT — Configure output triggers:
LINE — Select the output trigger link line (1 to 6).
VMC — Enable (ON) or disable (OFF) the VMC (voltmeter complete) output trigger.
DELAY — Configure the trigger delay of the trigger layer:
MAN — Manually set the delay (0 to 999.9998 seconds).
AUTO — Selects auto delay. The delay is set according to function/range (see Table
9-1).
COUNT — Set a finite (FIN) trigger count (1 to 2500), or select infinite (INF) trigger
count.
Input trigger and output trigger cannot share the same trigger link line. Defaults set
line 1 as the input and line 2 as the output.
DELAY — Configure the trigger delay of the trigger layer:
MAN — Manually set the delay (0 to 999.9998 seconds).
AUTO — Selects auto delay. The delay is set according to function/range (see Table
9-1).