17-8
SCPI Reference Tables
Table 17-4 (cont.)
SENSe command summary
Command
:LGRoup <name>
:LGRoup?
:AUTO?
:ADIScharge
:LEVel <NRf>
:LEVel?
[:STATe] <b>
[:STATe]?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
[:STATe] <b>
[:STATe]?
:MEDian
:RANK <NRf>
:RANK?
[:STATe] <b>
[:STATe]?
Note: *RST default is ON and SYSTem:PRESet default is OFF.
Description
Specify autorange limit; HIGH or LOW.
Query upper limit for autorange.
Query state of autorange.
Path for auto discharge:
Set auto discharge level; -21e-6 to 21e-6.
Query auto discharge level.
Enable or disable auto discharge.
Query state of auto discharge.
Path to control the Digital Filter:
Select filter control; MOVing or REPeat.
Query filter control.
Specify filter count; 2 to 100.
Query filter count.
Enable or disable digital filter.
Query state of digital filter.
Path to control median filter:
Specify "n" for rank; 1 to 5 (rank = 2n+1).
Query rank.
Enable or disable median filter.
Query state of median filter.
Default
parameter
Ref
SCPI
HIGH
Sec 5
2e-6
OFF
Sec 6
REP
10
OFF
Sec 6
1
OFF