Keithley 6514 Instruction Manual page 219

System electrometer
Hide thumbs Also See for 6514:
Table of Contents

Advertisement

A) SYSTem:PRESet
Returns the instrument to states optimized for front panel operation. SYSTem:PRESet
defaults are listed in the SCPI tables in Section 17.
B) SYSTem:TIME:RESet
Resets the absolute timestamp to 0 seconds. The timestamp also resets when power is cycled,
or after the instrument is on for 99,999.99 seconds. The TRACe:TSTamp:FORMat command is
used to select the absolute timestamp. See Section 8 (Buffer) for details.
C) SYSTem:POSetup <name>
Parameters
RST
PRESet
SAVx
The *RST and SYSTem:DEFaults are listed in the SCPI tables in the Section 17. A setup is
saved in memory using the *SAV command. See Section 14 (Common Commands) for details.
D) SYSTem:VERSion
Read the version of the SCPI standard being used by Model 6514. Example response mes-
sage: 1996.0.
E) SYSTem:KEY <NRf>
Parameters
1
=
2
=
3
=
4
=
5
=
6
=
7
=
8
=
9
=
10 =
11 =
12 =
13 =
14 =
15 =
16 =
DISPlay, FORMat, and SYSTem
Power-up to *RST defaults
Power-up to SYSTem:PRESet defaults
Power-up to setup stored in memory (x = memory location 0,
1, 2, 3 or 4)
SHIFT key
V key
I key
Ω key
Q key
XFDBK key
ZCHECK key
ZCOR key
-------
-------
RANGE up arrow key
AUTO key
RANGE down arrow key
ENTER key
Cursor right arrow key
GRD key
17 =
LOCAL key
18 =
AVG key
19 =
MEDIAN key
20 =
REL key
21 =
LIMIT key
22 =
DIGIT key
23 =
RATE key
24 =
Cursor left arrow key
25 =
-------
26 =
STORE key
27 =
RECALL key
28 =
DELAY key
29 =
DAMP key
30 =
HALT key
31 =
TRIG key
32 =
EXIT key
16-9

Advertisement

Table of Contents
loading

Table of Contents