SCPI programming
Table 2-2
SCPI commands — autozero
Command
SYSTem
:AZERo
[:STATe] <b>
Programming example
The following command sequence will perform one zero corrected amps measurement:
SYST:AZER OFF
SYST:AZER ON
Connection fundamentals
The following provides important fundamental information on input connections to Model
6514. Typical connection drawings are included with the various measurement procedures pro-
vided in subsequent sections of this manual.
Input connector
The rear panel INPUT connector is a 3-lug female triax connector that will mate to a cable
terminated with a male triax connector.
Description
SYSTem Subsystem:
Enable or disable autozero.
' Disable autozero.
' Enable autozero.
Measurement Concepts
Default
ON
2-3