[SOURce:]VOLTage:PROTection[:LEVel] < value >|MIN|MAX
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Sets the over-voltage protection level. Units are in volts. If the output voltage exceeds the OVP level, the output is dis-
abled and the Questionable Condition status register OV bit is set.
Parameter
0 – 120% of voltage rating|MIN|MAX,
*RST 120% of rating
Sets the over-voltage protection to 60 V: VOLT:PROT 60
An over-voltage condition can be cleared with the OUTput:PROTection:CLEar command after the cause of the con-
l
dition has been removed.
Agilent N8900 Series Operating and Service Guide
Typical Return
<over-voltage level>