To exit the command menu press the Meter key to immediately return to meter mode, or press the Menu key to
l
return to the top level.
For a map of the front panel menu commands, refer to
Menu example - accessing over-voltage protection.
Press the Menu key to access the front panel command menu. The first line identifies the menu path. When the menu
is first accessed, the menu is at the top or root, and the path is empty. The second line indicates the commands that
are available at the present menu level. In this case, the top-level menu commands are shown, with the Output com-
mand highlighted. The third line indicates which commands are available under the Output command. If there are no
lower level commands, a brief description of the highlighted command is displayed.
Press the right arrow navigation key > to traverse the menu until the Protect command is highlighted. Press the
Select key to access the Protect commands.
Since the OVP command is already highlighted, press the Select key to access the OVP dialog.
Note that the default OVP setting for all models is 120% of the rated ouptut voltage. You can change the OVP setting
using the numeric entry keys. Then press Select. Press the Meter key to return to meter view.
Agilent N8900 Series Operating and Service Guide
Front Panel Menu
Reference.
Getting Started
43