CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept any calibration
commands. The first parameter specifies the state. The second optional parameter is the password.
Parameter
0|OFF|1|ON , default OFF
<password> a numeric value up to 15 digits
Disable calibration mode and unsecure calibration: CAL:STAT OFF,0
Enable calibration: CAL:STAT ON
<password> is optional to secure the instrument, but must be correct if provided.
CALibrate:VOLTage[:LOCal][:LEVel] < value >
Calibrates the local voltage programming and measurement. The value selects the range to calibrate.
Parameter
The maximum voltage of the range
being calibrated.
Calibrates the voltage of the 80 V range: CAL:VOLT 80
CALibrate:VOLTage:REMote[:LEVel] < value >
Calibrates the remote voltage programming and measurement. The value selects the range to calibrate.
Parameter
The maximum voltage of the range
being calibrated.
Calibrates the voltage of the 80 V range: CAL:VOLT:REM 80
Agilent N8900 Series Operating and Service Guide
Typical Return
0 or 1
(none)
Typical Return
(none)
Typical Return
(none)
Calibrate Commands
79