Agilent Technologies N5769A Service Note

100v, 15a, 1500w
Hide thumbs Also See for N5769A:

Advertisement

Quick Links

S E R V I C E
N5769A Power Supply, 100V, 15A, 1500W
Serial Numbers: All Units
During verification performance testing of this power supply it can have a
unexpected voltage level on its output as result of the verification procedure
ending with the maximum voltage setting and not instructing the user to set
voltage levels to zero at the end of the verification procedure
Parts Required:
P/N
NONE
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
[[]] ON SPECIFIED FAILURE
CATEGORY: X AGREEABLE TIME
LOCATION X CUSTOMER INSTALLABLE
CATEGORY: [[]] ON-SITE
[[]] SERVICE CENTER
[[]] CHANNEL PARTNER
AVAILABILITY:
ALWAYS
[[]] Calibration Required
X Calibration NOT Required
ADDITIONAL INFORMATION: Customer can view latest manual changes at www.agilent.com/find/n5700
© AGILENT TECHNOLOGIES, INC. 2013
PRINTED IN U.S.A.
July 17, 2013
Rev. 21
Description
(active On-site contract required)
N5769A-03
N O T E
Qty.
STANDARDS
LABOR: 0.1 Hours
SERVICE
[[]] RETURN
INVENTORY: [[]] SCRAP
X SEE TEXT
NO CHARGE AVAILABLE UNTIL: EOS
PRODUCT LINE: SP
AUTHOR: PF
Supersedes:
NONE
USED
[[]] RETURN
PARTS:
[[]] SCRAP
X SEE TEXT
Page 1 of 2

Advertisement

loading

Summary of Contents for Agilent Technologies N5769A

  • Page 1 N O T E Supersedes: NONE N5769A Power Supply, 100V, 15A, 1500W Serial Numbers: All Units During verification performance testing of this power supply it can have a unexpected voltage level on its output as result of the verification procedure...
  • Page 2 Page 2 of 2 N5769A-03 Situation: When performing verification of power supply the last performance test instructs user to set the voltage setting to maximum rating. If power supply is set to auto restart mode it restores the operating setting that was set when it was last turned off. When this occurs the voltage or current will be set to maximum rating the next time it is turned on.